12
PM-1795-01-06/2018
• Do not use with a defibrillator.
• Do not install the unit in the following locations:
- Locations subject to vibration such as ambulances and emergency helicopters.
- A location where there is gas or flame.
- A location where there is water or steam.
- A location where chemicals are stored.
• Do not use at extremely high temperature, high humidity, or high altitude. Use only
within the required ambient conditions.
• Do not subject the unit to intense shock.
• Do not place heavy objects on the AC adapter cable, or allow the unit to sit on the
cord.
• Clinical testing has not been conducted on newborn infants and pregnant women.
Do not use on newborn infants and pregnant women.
• Do not plug in or unplug the AC adapter with wet hands.
• During measurement, make sure that no mobile phone or any other electrical
devices that emit electromagnetic fields is within 30 cm of this monitor. This may
result in incorrect operation of the monitor and/or cause an inaccurate reading.
Cautions
• Do not install the unit in the following locations:
- Locations with dust, salt, or sulfur.
- Locations directly exposed to sunlight for extended periods of time (in particular, do
not leave in direct sunlight or near a source of ultraviolet light for extended periods,
as ultraviolet light will cause deterioration of the LCD).
- Locations subject to vibration or shock.
- Near heaters.
• Do not use in a location with noise-emitting equipment such as a room with MRI, CT,
X-ray, or HF surgical equipment or an operating room. Noise from the equipment
may interfere with the operation of the unit.
Before use / during use
Warnings
• The unit complies with the EMC standard (EN60601-1-2). As such, it can be used
simultaneously with multiple medical instruments. However, if instruments that
generate noise such as an electric scalpel or a microwave therapy device are near
the unit, check the operation of the unit during and after use of these instruments.
• If an error occurs or a measurement result is questionable, check the vital signs of
the patient by auscultation or palpation. Avoid relying solely on the measurement
results of the unit when judging the patient’s condition.
• Only trained healthcare providers should use this device. Do not allow patients to
operate this device.
• Properly connect the connectors and AC adapter cable.
• Do not place objects or liquids on top of this unit.
• Check the following before using the unit:
- Make sure the AC adapter cable is not damaged (wires are not exposed or broken),
and the connections are firm.