10
3. Using the Unit
Important:
If your systolic or diastolic pressure is outside the standard range, the blood
pressure colour indicator will light in orange when the measurement result is
displayed. If they are within the standard range, the blood pressure colour
indicator will light in green.
Recent research suggests that the following values can be used as a guide to
high blood pressure for measurements taken at home.
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria, please refer to Chapter 7.
Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement. The irregular heartbeat algorithm
automatically determines if the measurement is usable or needs to be repeated. If the
measurement results are affected by irregular heartbeats but the result is valid, the result is
shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause
the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( )
is shown after you have taken a measurement, repeat the measurement. If the irregular
heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
If you move during measurement, the movement error symbol ( ) will appear on the
display. Keep still and repeat the measurement.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HEM-7320 and is the answer not in the manual?

Omron HEM-7320 Specifications

General IconGeneral
ModelHEM-7320
CategoryBlood Pressure Monitor
Irregular Heartbeat DetectionYes
Body Movement DetectionYes
Measurement Range Pressure0-299 mmHg
Accuracy Pressure±3 mmHg
Accuracy Pulse±5% of reading
InflationAutomatic by electric pump
DeflationAutomatic pressure release valve
Measurement MethodOscillometric method
Cuff Size22 – 42 cm
DisplayLCD
Power Source4 AA batteries or AC adapter
Device TypeUpper Arm
Measurement Range Pulse40-180 beats/min
WeightApprox. 380g without batteries