4. Troubleshooting and Maintenance
EN
Air plug disconnected.
Insert the plug securely.
Refer to section 3.1.
Arm cuff not applied correctly.
Apply the arm cuff correctly.
Refer to section 3.1.
Air is leaking from the arm cuff.
Replace the cuff with the new one.
Refer to Chapter 5.
Movement during measurement and the arm
cuff has not been inflated sufficiently.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to section 3.3.
If “E2” appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above your
previous measurement result.
Refer to section 3.3.
The arm cuff was inflated above 299 mmHg
when inflating the cuff manually.
Do not inflate the cuff above 299 mmHg.
Refer to section 3.3.
Movement during measurement.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to section 3.3.
Clothing is interfering with the arm cuff.
Remove any clothing interfering with the arm
cuff.
Refer to section 3.1.
Device error. Contact your local OMRON representative.
Error Display Cause Remedy

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HEM-7320 and is the answer not in the manual?

Omron HEM-7320 Specifications

General IconGeneral
ModelHEM-7320
CategoryBlood Pressure Monitor
Irregular Heartbeat DetectionYes
Body Movement DetectionYes
Measurement Range Pressure0-299 mmHg
Accuracy Pressure±3 mmHg
Accuracy Pulse±5% of reading
InflationAutomatic by electric pump
DeflationAutomatic pressure release valve
Measurement MethodOscillometric method
Cuff Size22 – 42 cm
DisplayLCD
Power Source4 AA batteries or AC adapter
Device TypeUpper Arm
Measurement Range Pulse40-180 beats/min
WeightApprox. 380g without batteries