27
LIST OF ERROR CODES
Error code Explanation
Inflation error
When the pressure does not exceed
12 mmHg within the set time after the
start of inflation
When the inflation does not reach the set
cuff pressure within the specified time
after the start of inflation
Deflation error
When the deflation speed is too fast during
the measurement
When the deflation speed is too slow
during the measurement
When the measurement does not finish
within the specified time after starting the
measurement
Overpressure error
The cuff pressure exceeded 299 mmHg.
Insufficient inflation error
Blood pressure could not be measured
due to insufficient inflation level.
Indeterminable blood pressure error
Blood pressure could not be measured
even when the cuff pressure reached the
specified pressure.
Low pulse level error
Pulse wave was too small.
Blood pressure error
Relationship between systolic and dias-
tolic pressures was abnormal.
Pulse rate error
Pulse rate did not stay within the range of
30 to 199 beats/min.
Device error
Main unit malfunction.
How to correct
Confirm that the air tube connecting the cuff and
the main unit is connected securely.
Confirm that the air flow in the air tube
connecting the cuff and the main unit isnt
being restricted.
Confirm that the cuff is wrapped correctly (refer
to pages 13 and 14).
Check bladder for leaks and, if necessary,
replace the bladder with new one (option).
Confirm that air flow in the air tube connecting
the cuff and the main unit isnt being restricted.
If the measurement is made by setting the P-SET
to AUTO, ask the patient not to move during
the inflation.
Confirm that the P-SET is securely set to
AUTO. Turn the Knob counterclockwise as far
as it goes until you can hear a click sound.
If the measurement is made by manual inflation
level setting, set the value to 30 to 40 mmHg
higher.
Confirm that the cuff is wrapped correctly (refer
to pages 13 and 14).
Confirm that the cuff is wrapped correctly (refer
to pages 13 and 14).
Ask the patient not to move during the measure-
ment.
Check the patient for arrhythmia.
Contact Omron Healthcares Cusomer Service
toll-free at 1-877-216-1336.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 27

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron IntelliSense HEM-907XL and is the answer not in the manual?

Omron IntelliSense HEM-907XL Specifications

General IconGeneral
BrandOmron
ModelIntelliSense HEM-907XL
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals