EasyManua.ls Logo

Optek TF16 - Maintenance; Preventive Maintenance

Optek TF16
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
- 24 -
optek-Manual--1004-2006-02--TF16-US-2011-06-20
www.optek.com
9 Maintenance
9.1 Preventive maintenance
Note!
The used detector is not subjected to measurable aging when used properly.
Tab. 11 Preventive maintenance
Component Maintenance activity Maintenance interval Information
Wetted parts
inspection with regard
to leakages
as part of standard
installation mainte-
nance
Possible damaged sealing faces
of the sensor body can lead to
leakages (see instruction manual
of the sensor body).
Lamp
module
exchange 1 - 2 years
Operation of the lamp below its
nominal voltage (4.8 V DC
instead of 5.0 V DC) enhances
lamp life. Strong vibrations, high
temperatures or frequent on and
off switching of the system can
have negative effects on the ser-
vice life. Statistical service life
amounts to 3 years, for UV
lamps to 1–2 years.