Opti-Sciences, Inc. CCM-200 plus operator's manual
18
Reconnect the CCM to the PC. Go to My Computer, find the CCM200 drive, and right click the
drive. Left click Format, and then left click Start. This will reformat the drive. All data that was
on the instrument will be lost. With earlier versions of software, it is important to either push the
cancel button or disconnect the USB cable before turning off the instrument.
Message during app upgrade:
" CCM200p LOADER"
" Updating Application "
" Please Wait!"
This message is displayed during an application upgrade.
" ! nvRAM Init !"
This message will sometimes follow a software update. It may also appear after the unit is
formatted by the PC. It indicates that the system defaults are being restored. It should not appear
every time the unit is turned on. That would indicate a uSD card fault.
Test errors:
"Custom file not found!"
"CCI units will be used"
Fix:
This message will appear if the "Use custom measurementunits [Yes]" setting is selected. It
means that the lookup table for this feature is not present. Simply turn feature off to clear the
message.