Opti-Sciences, Inc. CCM-200 plus operator's manual
19
"Warning! Average must "
"be>=10 pts for std dev"
"mode. A setting of 10 "
"points will be used."
Fix:
This message will appear for the standard deviation test mode when the number of samples is
less than 10 points. It must be 10 or greater.
" !! Warning !!"
" Drive is Full"
Fix:
This will appear if the devices memory is full. Delete some unused data to remedy this problem.
"Sample too dense"
Fix:
This message warns that the amount of signal absorbed by the sample is too great for reliable
measurements. Try locating to a different part of the sample.
" Hold arm down "
" for whole test!"
Fix:
This message appears if the sample chamber is not closed for the duration
of the test.