Applying Control Solution to
the Electrode
1
Turn the control solution
bottle upside down three
to four times to mix
the solution.
2
Remove the cap.
3
Apply a drop of control
solution to the electrode in
the area shown here. The
control solution is drawn
into the electrode.
4
Continue to touch the
control solution to the
electrode until the monitor
begins the test.
The monitor begins the
test when:
You hear the beeper, if the
beeper is ON.
The display window shows
the status bar.
Then the display window
shows the countdown.
Glucose
ß-Ketone
61
Status
bar
Countdown
Control Solution
Testing
133-221 Manual,G3b,Aus 3/17/05 11:15 AM Page 64