Glucose
ß-Ketone
Note: Do not
remove the
electrode from the monitor or
disturb it during the countdown.
Important: If the countdown
does not start:
What It Means:
You might not have applied
enough control solution to
the electrode.
What to Do:
Apply a second drop of
contr
ol solution to the
electrode within 30 seconds
of the first drop.
If the countdown still does
not start or if more than 30
seconds have passed, discard
the electrode, turn off your
monitor, and try again with a
new electrode.
5
At the end of
the countdown:
If the beeper is ON, listen for
the beeper.
The control result shows on
the display window.
Control
Solution Testing