12
Operation
Locating and dealing with environmental disturbances
During testing outlined on page 11, if the alarm does not sound after
20 seconds, eliminate all environmental disturbances by actions
such as changing ventilation (the direction of the airow of a fan or
air conditioning vent), removing cot mobile, closing windows, and
repositioning the cot etc.
Babysense2 cannot be used for two infants in one cot. Never allow
animals in the cot. Do not operate vaporizer or ventilator directly towards
the cot.
In the event of alarm
In the event of alarm commence rst aid, and seek the assistance of
emergency services and/or medical advice.
IMPORTANT: It is strongly advised that every caregiver to your
baby learn how to administer rst aid.
False alarms
Isolated cases of false alarms are possible. If this occurs, check if this is
caused by:
• a faulty connection between the sensor pad(s) and the control unit;
• connection of only one sensor pad when your baby can move off the
immediate area;
• connection of two sensor pads far apart from each other when baby is
not crawling yet; or
• failure of the sensor pads to be directly against the mattress.
• The compositions of some mattresses may not conduct vibrations
associated with babies breathing movement and may not be suitable
with Babaysense2. Do not use Babysense2 with a memory foam
mattress.
• In case of false alarms, kindly refer to troubleshooting guide.
Please contact Oricom Customer Support team with any product related
questions.
Please note
Babysense2 monitors and alerts with respect to breathing
movement rates, but cannot prevent cessation of breathing. The
responsibility for applying rst aid rests on you.