6. COMMANDS AND RESPONSES
23
%000038080<CR>
Device preset already exceeded and High Voltage not enabled. The
command was ignored.
%000054078<CR>
Device preset already exceeded, High Voltage not enabled and
amplifier was not pole-zeroed. The command was ignored.
Power-Up Alert: Power-up just occurred and the selftest results are:
%001000070<CR>
All power-up selftest passed.
%003000072<CR>
Battery backed-up data lost.
Command Syntax Errors:
%129001082<CR>
Invalid command verb.
%129002083<CR>
Invalid command noun.
%129003084<CR>
Invalid command verb and noun.
%129004085<CR>
Invalid command modifier.
%129005086<CR>
Invalid command verb and modifier.
%129006087<CR>
Invalid command noun and modifier.
%129007088<CR>
Invalid command verb, noun, and modifier.
%129132087<CR>
Invalid command (verb, noun, and modifier valid but not together).
Communication Errors:
%130001074<CR>
Serial line buffer overrun.
%130002075<CR>
Serial line parity error.
%130003076<CR>
Serial line buffer overrun and parity error.
%130004077<CR>
Serial line framing error.
%130005078<CR>
Serial line buffer overrun and framing error.
%130006079<CR>
Serial line parity error and framing error.
%130007080<CR>
Serial line buffer overrun, parity error, and framing error.
%130008081<CR>
Serial line break detected.
%130128084<CR>
Command checksum incorrect (only when optional checksum
provided).
%130129085<CR>
Command (or WRITE handshake) record too long.
%130131078<CR>
WRITE command aborted by HA handshake.
%130132079<CR>
WRITE command aborted by timeout.
%130133080<CR>
WRITE command aborted by invalid handshake.
Execution Errors:
%131128085<CR>
Invalid 1st command parameter.
%131129086<CR>
Invalid 2nd command parameter.
%131130078<CR>
Invalid 3rd command parameter.
%131132080<CR>
Invalid number of command parameters.
%131134082<CR>
Invalid device or segment selected.

Questions and Answers: