DSPEC
®
Digital Gamma-Ray Spectrometer
44
from 001031, month as a 3-digit integer number from 001012, and year as a 3-digit integer
number from 000099. See also SET_DATE_START.
Responses:
$N001001088052<CR>
Date reported as Jan 1, 1988.
... ...
$N031012099059<CR>
Date reported as Dec 31, 1999.
$N001001000036<CR>
Date reported as Jan 1, 2000.
... ...
$N031012087056<CR>
Date reported as Dec 31, 2087.
SHOW_DATE_START
Returns the day, month and year of the acquisition start date in the form dddmmmyyy. The
day is returned as a 3-digit integer number from 001031, month as a 3-digit integer number
from 001012, and year as a 3-digit integer number from 000099. See also
SET_DATE_START.
Responses:
$N001001088052<CR>
Date reported as Jan 1, 1988.
... ...
$N031012099059<CR>
Date reported as Dec 31, 1999.
$N001001000036<CR>
Date reported as Jan 1, 2000.
... ...
$N031012087056<CR>
Date reported as Dec 31, 2087.
SHOW_DCREST
Shows the DC restorer setting. See also SET_DCREST.
Responses:
DCRE002000
SHOW_DEBUG
Shows the debug level. See also SET_DEBUG.
Responses:
$A003248
Returns the Debug state as $Axxxccc, where xxx is the
Debug level and ccc is the checksum.
SHOW_DELAY_COLLECTION
Shows the Delay Collection setting. See also SET_DELAY_COLLECTION.
Responses:
DEL_COLL 0000000000001.6