EasyManua.ls Logo

ORTEC DSPEC - Page 65

ORTEC DSPEC
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. COMMANDS AND RESPONSES
59
SHOW_ZERO_PRESET
Reports the Gauss mode stabilization preset for the zero peak. The preset represents the
minimum number of incremental counts that must be collected in any one channel of the zero
peak before the zero offset is evaluated by the stabilizer and potentially adjusted. See also
SET_ZERO_PRESET and CLEAR_PRESETS.
Responses:
$G0000000010076<CR>
Zero preset currently 10 counts (minimum).
$G0000000011076<CR>
Zero preset currently 11 counts.
... ...
$G2147483647121<CR>
Zero preset currently 2147483647 counts (maximum).
SHOW_ZERO_STABILIZATION
Reports the state of zero peak stabilization. See also ENABLE_ZERO_STABILIZATION
and DISABLE_ZERO_STABILIZATION.
Responses:
$IT<CR>
Zero stabilization is currently enabled.
$IF<CR>
Zero stabilization is currently disabled.
SHOW_ZERO_WIDTH
Reports the current width for the stabilizer zero peak. See also SET_ZERO_WIDTH,
SET_ZERO_CHANNEL, and SHOW_ZERO_CHANNEL.
Responses:
$C00001088<CR>
Zero width has not been set.
$C00003089<CR>
Zero width is 3 channels (lowest possible width).
... ...
$C00256100<CR>
Zero width is 256 channels (highest possible width in Gauss
mode).
... ...
$C16383108<CR>
Zero width is 16383 channels (highest possible width in point
mode with zero channel set to 8192).
START
[seg-mask]
Starts the acquisition of spectral data. The optional segment mask is provided for
compatibility with other MCBs and may be any value from 0 to 65535 but is ignored by the
DSPEC.
Execution Warnings:
%000004073<CR>
No segment selected (occurs with other warnings).
%000005074<CR>
The acquisition is already started (no changes made).
%000006075<CR>
A preset was exceeded (acquisition was not started).
%000008073<CR>
Unit not optimized since initialization.
%000016076<CR>
Amplifier not PZd since initialization.
%000032074<CR>
High voltage is disabled.

Related product manuals