ID: 1588 / ver. 205
15
Technical specifications
Standards: EN 60601-1: 1996
EN 60601-1-2:2001
EN 60645-5:2005
Classification: Group 1, class A EN 60601-1-2:2002
Medical CE- mark: Inmedico A/S is approved for medical CE marking, by DGM.
Identification number 0543
Tympanometer:
Test frequency: 226 Hz
Tolerance frequency: ± 1%
Test Level: 85dB SPL in 2ml ± 3dB
Pressure: +200 to -300 daPa
Volume range: 0.1 to 6.0 ml
Reflex measurements:
Test frequencies: 500 Hz, 1 kHz, 2 kHz, 3 kHz, 4 kHz ± 2%
Test method: ipsilateral
Test Level: 105dB SPL in 2ml ± 3dB
Supply: 230 VAC / 12 VDC ±10 %
Dimension: 225 mm x 180 mm x 55 mm
Weight: Approx. 1000 g
Warm-up time: < 10 minutes.
Operating temperature: +15 to +35 degree C.
Humidity: 30 % to 90 %
Surroundings pressure: 98 to 104 kPa
Calibration cavities: 0.5 ml, 2.0 ml and 5.0 ml, the volume tolerances shall be ± 2 % or 0.05 cm
3
,
whichever is greater.
Included parts: 1 PSU, 1 user manual, 1 calibration certificate, 1 calibration cavity set and 1 probe
with 6 sizes of tips.
Accessories: Carrying bag, thermo printer.