Technical data
Measuring range: Measuring range - differential pressure: -0.05 kPa to 200 kPa
Max. static excess pressure: 2000 kPa
Measuring range – temperature: -20°C to +120°C
Resolution: Differential pressure: 0.01 kPa
Flow: 0.0001 l/s
Temperature: 0.1°C
Accuracy: Differential pressure: up to 10 kPa ± 0.1 kPa
10 to 200 kPa 1% of measured value
Flow: 0.01 l/s
Temperature ± 1°C
Ambient temperature: Working temperature: 0°C to 40°C
Storage: -20°C to +60°C
Attention: In case of risk of frost, no water may remain in the
measuring head!
In this case, the measuring head and the measuring hoses have to
be drained completely.
Humidity: Max. relative humidity of 90% non-condensing
Protection class: Body IP52
Keyboard IP54
Dimensions/weight: Flow meter: 160 x 63 x 40 mm, weight 470 g
Measuring head: 130 x 70 x 70 mm, weight 1240 g
Display: LCD display with backlighting
Power supply: 4 rechargeable NiMh batteries
or by using enclosed power pack 230 V AC 50/60 Hz
or by using 4 commercial Mignon batteries 1.5 V
Attention: In case of battery operation, the power pack must not be
connected!
Interface: USB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the oventrop OV-DMC 2 and is the answer not in the manual?

oventrop OV-DMC 2 Specifications

General IconGeneral
Brandoventrop
ModelOV-DMC 2
CategoryMeasuring Instruments
LanguageEnglish