Instructions for use
remOVE DC Impulse 400.01
English; Version 08 | 2019-04-24
Appendix 1
Inspection sheet
Annual technical safety control
Make sure the DC Cord is correctly plugged into the generator.
Do not press the foot-activated switch while testing the device.
Ensure that country-specific regulations and requirements, which include the generally accepted technical stand-
ards, as well as valid regulations for occupational safety and accident prevention, will be implemented and com-
plied with during setup, operation and maintenance.
Device: remOVE DC Impulse
Safety classification: II
Article no.: 400.01
Type of applied part: BF
Serial no.:
Equipment:
Inspection before initial start-up (refer-
ence value)
Repeated inspection
Inspection after maintenance
Foot-activated switch
DC Cord
Power cord
Test Instrument for electrical tests
1. Electrical tests (The tests must be executed
when the device is fully charged)
Result
1.1 Are all cable insulations intact?
Insulation must be intact
passed failed
annex, where appropriate
1.2 Insulation resistance according to the standard
EN 62353:2014 figure 4
passed failed
annex, where appropriate
1.3 Direct measurement for the device leakage
current according to the standard EN
62353:2014 figure 7
passed failed
annex, where appropriate
1.4 Direct measurement of the leakage current of
the applied part according to the standard EN
62353:2014 figure 10
passed failed
annex, where appropriate