English; Version 08 | 2019-04-24
2.1 Are all components available?
Components must be complete
☐ passed ☐ failed
☐ annex, where appropriate
2.2 Do all LEDs light up correctly? (On/Off display,
status display)
All LEDs must light up according to the in-
structions for use.
☐ passed ☐ failed
☐ annex, where appropriate
2.3 Is the successful completion of the charging
cycle indicated by three short acoustic signals?
Completion of charging cycle must be indi-
cated by three short acoustic signals
☐ passed ☐ failed
☐ annex, where appropriate
Overall evaluation:
☐ No Safety or functional defects were detected.
☐ No direct risk, the detected defects can be fixed shortly.
☐ The device has to be withdrawn from use until the defects are corrected.
☐ The device does not meet the requirements – Modifications/Replacement of com-
ponents/removal from service is recommended.
Contact the manufacturer if there is any deviation, abnormality and/or an indica-
tion of error.
The next inspection should be carried out after 12 months.