10
6.8.3 1 Point Calibration
Steps:
1. Select Calibration from the main menu.
2. Select Enter Password from the Calibration menu. Enter “0000” if you have not set your
own password yet.
3. Select 1 Point Cal. from the sub-menu.
4. Select Expected Volume 1. By default, 10% of the maximum volume will be displayed.
You can change this volume depending on your preferences using the Up/Down arrow
keys.
5. Enter Delivered Volume 1.
6. Select “Yes” to save your changes. The display will read Calibration done. Select “No” to
exit without saving.
6.8.4 Restore Factory Calibration
This feature can be used to restore the calibration settings to the factory default. All user
generated changes will be cleared and the factory calibration will be restored.
Steps:
1. Select Calibration from the main menu.
2. Select Enter Password from the Calibration menu. Enter “0000” if you have not set your
own password yet.
3. Select Restore Fact. Cal.
4. Select “Yes” using the Up/Down arrow keys to restore the factory calibration settings.
“Factory Calibration Restored” will be displayed on the screen for a few seconds. You
will then be directed to the calibration menu.
5. Select “No” using the Up/Down arrow keys to keep the user defined settings.
Special Note: Whenever a user makes any changes to calibration, the screen will display
“Calibration User”. When a user restores the factory calibration settings, the screen will
display “Calibration Fact.”.
6.8.5 Last Calibration Date
Here, the most recent calibration date is logged. Whenever you calibrate your pipette, you
can enter the calibration date here.
Steps:
1. Select Calibration from the main menu.
2. Select Enter Password from the Calibration menu. Enter “0000” if you have not set your
own password.
3. Select” Last Cal. Date” using the Up/Down arrow keys.
4. Enter the date in DD/MM/YY format. You will then be directed to the calibration menu.