10. Autoclavability
The lower part of the pipette, tip cone, ejector collar, piston and O-ring are
autoclavable parts. Please autoclave under the conditions below.
Temperature
:
121°C / 25Fahrenheit
Pressure
:
1 bar / 15 psi (Pounds)
Time
:
25 minutes or less
After autoclaving, allow parts to cool down for 8 hours in normal condition
before reassembling the unit. Do not autoclave the upper electronic portion
of the pipette.