4. iCELLis 500+ Control System Environment Requirements
WARNING! Risk of injury to personnel and/or damage to equipment!
The iCELLis 500+ control system is not ATEX certified. The system may not be operated in
potentially explosive atmosphere.
DANGER! Risk of injury to personnel and/or damage to equipment!
Before installing the iCELLis 500+ control system, verify that the floor can support the weight of the
system, the ancillary equipment and their content (such as cell culture media).
ALERT! Risk of damage to equipment!
This equipment has been designed for indoor use only and a maximum ambient environmental
The dimensions and weight of the iCELLis 500+ control system can be found in Table 1 below.
Table 1: iCELLis 500+ control system dimensions and weight
1850 mm (excluding visual alarm and tubing holders)
2122 mm (with visual alarm and tubing holders)
955 mm (excluding optional high flow pump)
1038 mm (with optional high flow pump)
Dry Weight
(without bioreactor)
Maximum weight when loaded
(with bioreactor full of media and
double jacket full of water)
Environmental working conditions of the iCELLis 500+ control system are detailed in Table 2 below.
Table 2: iCELLis 500+ control system environmental working conditions
Environmental Working Conditions
Ambient operating temperature
Relative operating humidity
The work area where the iCELLis 500+ control system is installed and operated must be leveled and be
able to accommodate the footprint and weight of the iCELLis 500+ control system with a 650 mm
minimum clearance on each side of the system to enable access to the controller buttons (such as
emergency stop buttons) and other inlets/sockets.