EasyManua.ls Logo

Pall iCELLis 500+ - Temperature Control Unit Specifications

Pall iCELLis 500+
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 42 of 48
WT_02 Weight bioreactor 0 ... 180,0
+/- 1.0% of
liquid weight
(agitation from 0
- 300 rpm)
+/- 3.0% of
liquid weight
(agitation from
300-450 rpm)
Kg
ST_01
Speed feed-in pump
(Watson Marlow 520R pump head)
0 ... 360 +/- 2.0 rpm
ST_02
Speed base pump
(Watson Marlow 313DW pump head)
0 ... 360 +/- 2.0 rpm
ST_03
Speed inoculum pump
(Watson Marlow 313DW pump head)
0 ... 360 +/- 2.0 rpm
ST_04
Speed feed-out pump
(Watson Marlow 520R pump head)
0 ... 360 +/- 2.0 rpm
ST_05
Speed sampling pump
(Watson Marlow 313DW pump head)
0 ... 360 +/- 2.0 rpm
ST_06
Speed agitation system
0 ... 450
+/- 5.0
rpm
ST_07
Speed fill pump (optional)
(Watson Marlow 630BpN)
0 ... 160 +/- 2.0 rpm
ST_08
Speed drain pump (optional)
(Watson Marlow 630BpN)
0 ... 160 +/- 2.0 rpm
6.3 Temperature Control Unit Specifications
Description
LAUDA PRO RP 240 EC
230V 50Hz
LAUDA PRO RP 240 EC
120V 60Hz
Working temperature range (°C)
-40 ... +200
-40 ... +200
Temperature stability (°C)
±0.05
±0.05
Setting / display resolution
0.01 °C
0.01 °C
Temperature display
TFT touchscreen
TFT touchscreen
Electric current consumption (A)
16
16
Heating capacity (kW)
2.5
1.8
Pump capacity flow rate (L/min)
22
22
Pump capacity flow pressure (bar)
0.7
0.7
Pump capacity flow suction (bar)
0.4
0.4
Process volume min (L)
2.4…4.4
2.4…4.4
Digital interfaces
Profibus
Profibus
Ambient temperature (ºC)
5...40
5...40