EasyManua.ls Logo

Pall iCELLis 500+ - Page 33

Pall iCELLis 500+
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 33 of 48
Figure 25: Connection of PROFIBUS cable between TCU and iCELLis 500+ control system
a) Front of TCU, including Interface panel, b) Side of TCU with Interface panel,
c) Connections at the rear of iCELLis 500+ control system.
Follow Figure 26 to assemble and install the TCU.
Remove the TCU unit from the crate and install it on a flat surface (Figure 26a-b). Place the command
touch panel in its designated place on the TCU and connect its cable to the right side of the TCU as
shown in (Figure 26c-d).
Take out the Profibus cable from crate and connect one end of the cable to right side of the TCU and
other end to back of the iCELLis 500+ control system, tighten the connector on the TCU using a flat
screwdriver, please refer to Figure 26e-f.
Confirm the TCU power switch is in the OFF position and connect the power cable to the appropriate
voltage supply (230Vac - 50Hz or 120Vac - 60Hz). Refer to the user manual of the TCU for power
options. Refer to Figure 26h-i.
Locate the barbed connectors from the crate and remove the nuts from the back side of the TCU
IN/OUT lines. Refer to Figure 26j-k. Refer to Figure 26l-o to put nuts on the connectors and connect the
combination to the IN/OUT line of TCU.