Page 40 of 48
Double jacket over-pressure
protection
Interlock using disposable pressure sensor for measurement of the double
jacket internal pressure
2 emergency stop buttons, 1 on the front and 1 on the left side of the
iCELLis 500+ control system Top Assembly.
Power failure data protection
Integrated UPS for the safe shut down of the computer only. A power
supply UPS is highly recommended
Wonderware
♦
ArchestrA system platform
2 x SSD of 240 Go (RAID1 for redundant data storage)
electronic signatures
Compatible with FDA 21 CFR Part 11 and Eudralex Annex 11
Developed and tested in accordance with GAMP5 guidelines
Data export/Network
communication
Integrated OPC server for real-time data monitoring. Access to SQL
database for historical data storage
19 in. touch panel & USB interface
Available via Windows
♦
Remote Desktop Protocol
Recommended cleaning solutions for wiping exposed surfaces of
the hardware:
• Water
• 70% (v/v) ethanol in water
•
Recommended preventative
maintenance interval
Maximum relative humidity in
storage
Relative humidity during
operation
10 – 70% RH, non-condensing