Panasonic MCO-170AICUVH Series - STD GAS AUTO CALIBRATION KIT (OPTION); Connecting Standard Gas Cylinder

Panasonic MCO-170AICUVH Series
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74
STD GAS AUTO CALIBRATION KIT (OPTION)
When a STD gas auto calibration kit MCO-SG is installed, by connecting standard gas cylinder for
calibration, it is possible to calibrate CO
2
density manually.
1. Connect a standard gas cylinder to connecting
port for standard gas cylinder on lower right side of
the CO
2
incubator. Since a Standard gas cylinder is
used as a standard of exact density during CO
2
density calibration, prepare a standard gas cylinder
that is same as the set CO
2
density.
Note: There is not the problem to remain
connected standard gas cylinder after finished CO
2
gas density calibration.
2. Press Menu key to lead the Menu screen.
3. Press Tools #1 key to lead the Tools #1 screen.
4. Press STD Gas Setting key to lead the STD Gas
Setting screen.
Lower right side
Connecting port for standard gas cylinder