EasyManua.ls Logo

Panasonic MCO-170AICUVH Series - PERFORMANCE; Performance Characteristics

Panasonic MCO-170AICUVH Series
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
91
PERFORMANCE
Product name
CO
2
Incubator MCO-170AICUVH
MCO-170AICUV
MCO-170AIC
Model No.
MCO-170AICUVH-PE
MCO-170AICUV-PE
MCO-170AIC-PE
Temperature control range
Ambient temperature+5
o
C to 50
o
C* (ambient temperature: 5
o
C to 35
o
C)
Temperature distribution
±0.25
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
Temperature variation
±0.1
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
CO
2
control range 0 % to 20 %
CO
2
variation
±0.15 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
Chamber humidity 95 %R.H.±5 %R.H.
Applicable environment
condition
Temperature: 5
o
C to 35
o
C, Humidity: 80 %R.H. max.
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15
o
C)
Noise level 29 dB (A scale)
Power consumption
Max. 379 W
Heat emission
Max. 1,070 kJ/h
Rated voltage, frequency
AC 220 V-240 V, 50 Hz
Amperage
Max. 1.8 A
*When ambient temperature is 25
o
C, temperature control range: 30
o
C
50
o
C. Regardless of ambient
temperature, the maximum of temperature control range is always 50
o
C.
Note: The unit with CE mark complies with EC directives.
Based on our measuring method.