DECONTAMINATION
9. Set the humidifying pan cover, the humidifying pan removed in the procedure 1 (Fig. 3).
Note: H
2
O
2
decontamination can be performed only for the chamber and inner attachments with standard
specifications, not for any other objects.
10. Make sure that the duct, the fan and the H
2
O
2
generator are securely attached (refer to Fig. 1 to 3 on
page 97). After that, close the gastight split doors, the inner door and the outer door.
11. Press H
2
O
2
key for 3 seconds to open Setting
Position window.
Note:
When the model does not have H
2
O
2
decontamination function, H
2
O
2
key is not
displayed on the Top screen.
When key lock is ON, Password input window is
opened and input of the release password of Key
Lock is required. Refer to page 41.
12. Finally make sure that the inner attachments
and the H
2
O
2
generator are attached correctly. After
confirm, press Next key to lead the H
2
O
2
Decontamination screen.
Fig. 3
Front
Humidifying
pan cover
Humidifying pan
H
2
O
2
generator