101
PERFORMANCE
Product name
O
2
/CO
2
Incubator MCO-170MUVH
MCO-170MUV
MCO-170M
Model number
MCO-170MUVH-PA
MCO-170MUV-PA
MCO-170M-PA
MCO-170MUVH-PK
MCO-170MUV-PK
MCO-170M-PK
Temperature control range
Ambient temperature+5
o
C to 50
o
C* (ambient temperature: 5
o
C to 35
o
C)
Temperature distribution
±0.25
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Temperature variation
±0.1
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
CO
2
control range 0 % to 20 %
CO
2
variation
±0.15 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
O
2
control range 1 % to 18 % and 22 % to 80 %
O
2
variation
±0.2 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Chamber humidity 95 %R.H.±5 %R.H.
Applicable environment
condition
Temperature: 5
o
C to 35
o
C, Humidity: 80 %R.H. max.
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15
o
C)
Noise level 25 dB (A scale)
Power consumption
Max. 370 W Max. 375 W
Heat emission
Max. 1030 kJ/h
Rated voltage, frequency
AC 110 V-120 V, 60 Hz AC 220 V, 60 Hz
Amperage
Max. 3.3 A Max. 1.8 A
*When set temperature is 37
o
C, ambient temperature must be 32
o
C or less. Regardless of ambient
temperature, the maximum of temperature control range is always 50
o
C.