Panasonic MCO-170M Series - Maintenance and Support; Routine Maintenance Schedule

Panasonic MCO-170M Series
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
89
8. On the STD Gas Calibration Step2 screen, press
OK key to lead the STD Gas Calibration Step3
screen.
9. On the STD Gas Calibration Step3 screen, gas
density calibration starts. Calibration go to Step5
(Procedure 10) automatically.
10. After completion of gas density calibration,
display leads the STD Gas Calibration Step5
screen. O
2
/CO
2
incubator returns to the normal
operation.
11. On the STD Gas Calibration Step5 screen,
press OK key to return to the Tools #1 screen. On
the Tools #1 screen, press Top key to return to the
Top screen.