Manual N
2
gas (O
2
gas) supply line A/B changeover
It is possible to change N
2
/O
2
gas supply line A and B manually anytime.
Example) Change N
2
/O
2
gas supply line A to B.
1. Press N
2
/O
2
gas supply line select key for a few
seconds.
2. N
2
/O
2
gas supply line A is changed to B.
Note: The behavior for the following case is shown in Table 3.
After the N
2
/O
2
gas supply line A is changed to B by N
2
gas (O
2
gas) automatic changer function, the N
2
/O
2
gas supply line B is changed to A manually without pressing the Buzzer key.
Table 3
Situation
N
2
gas N
2
/O
2
gas supply line indicator
Message
display field
Supply line Cylinder A Cylinder B A B
1
N
2
/O
2
gas supply line
A is changed to B
automatically.
B Empty Remaining
Reverse
video
and blink
Light on
Err02:
N2/O2
Gas Empty
(and buzzer)
2
Without pressing the
Buzzer key,
long-pressed N
2
/O
2
gas supply line select
key.
A Empty Remaining
Blink Light off
Err02:
N2/O2
Gas Empty
(and buzzer)