- 18 -
5 TROUBLESHOOTING
Table: Faults
In the event of faults that ...
In the event of faults that are not listed in this chapter, or if the
suggested procedure does not correct the fault, contact the
PARI GmbH Service Center [See also: 7.4 Contact].
Fault
Possible cause
Procedure
The compres-
sor does not
start up.
The power cord is not
plugged into the device
properly.
Check that the power
cord is seated correctly
in the compressor power
cord socket.
The power plug is not
plugged into the mains
socket properly.
Check that the power
plug is seated correctly
in the mains socket.
No medication
is coming out
of the
nebuliser.
The nebuliser nozzle is
blocked.
Clean the nebuliser.
The connection tubing is
not connected properly.
Check that both ends of
the tubing are seated
properly.
The connection tubing is
leaking.
Replace the connection
tubing.