- 21 -
en
(transmitters) and the PARI compressor, as recommended
below in accordance with the maximum output power of the
communication device.
Safety distance depending on transmitting frequency:
For transmitters whose rated power does not appear in the table
above, the safety distance can be calculated using the equation
associated with the respective column, where "P" is the rated
power of the transmitter in Watt (W) according to the information
supplied by the device manufacturer.
6.5 Ambient conditions for operation
Rated power
of the transmit-
ter [W]
150 kHz to
80 MHz
D = (3.5/3) ×
SQRT(P) [m]
80 MHz to
800 MHz
D = (3.5/3) ×
SQRT(P) [m]
800 MHz to
2.5 GHz
D = (7/3) ×
SQRT(P) [m]
0.01
0.2
0.2
0.3
0.1
0.4
0.4
0.8
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Ambient temperature
+10 °C to +30 °C
Relative humidity
30% to 75% (non-condensing)
Atmospheric pressure
700 hPa to 1,060 hPa