- 20 -
6.3 Information about electromagnetic
compatibility
Electrical medical equipment is subject to special precautionary
measures with regard to electromagnetic compatibility (EMC).
Such equipment must only be installed and operated in accord-
ance with EMC instructions.
Portable and mobile high-frequency communication devices can
disrupt electrical medical equipment. Using accessories, con-
verters and power cords other than those specified, with the ex-
ception of converters and power cords that the manufacturer of
the medical electrical device sells as spare parts for internal
components, can result in higher emission levels or lower the
resistance to interference of the device.
The device must not be placed directly beside or on top of other
devices for operation. If the medical electrical device must be
placed beside or on top of other devices to operate it, it should
be monitored constantly to ensure that it is operating properly in
the arrangement used.
Technical data on electromagnetic compatibility (EMC instruc-
tions) in table format are included with the compressor. It can
also be ordered from PARI GmbH or downloaded from the inter-
net at www.pari.de/en/products under "Technical Data" on the re-
spective product page.
6.4 Recommended safety distances
Recommended safety distances between portable and mobile
high-frequency (HF) telecommunication devices and the PARI
compressor:
The PARI compressor is intended for operation in an electro-
magnetic environment in which radiated HF transients are mon-
itored. The user can help to prevent electromagnetic
interference by observing the minimum safety distances be-
tween portable and mobile HF communication equipment