Accuracy
+/- 0.30 psi at -5 to 30 psi for 2 point calibrated
sensors (can be improved to -5 to 60 psi with 3
point calibration, also available)
Resolution
0.01 psi
Temperature Range
0 to 60 °C (32 to 140 °F).
Sensor Microchip
EPROM, Stores Sensor ID, and
Calibration Factor.
Sensor Connector & Cables
Lockable and Waterproof.
SciTemp:
Power
5 VDC provided by DINTemp
®
Module
Fluid Connections
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Wetted Materials
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Sensor Type
Thermistor, Epoxy coated, 2252
Temperature Range
-40 to +150° Celsius
Accuracy
+/- 0.10° Celsius in the 4.0 to 70° Celsius range.
Resolution
0.01 ° Celsius
Sensor Microchip
EPROM, Stores Sensor ID, and
Calibration Factor.
Sensor Connector & Cables
Lockable and Waterproof.