Specifications – Sensors
SciCon:
Power
5 VDC provided by DINCon
®
Module
Fluid Connections
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Wetted Materials
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Sensor Type
4-Electrode Conductivity Cell, gold plated pins.
Conductivity Range
1 μS/cm to 200 mS/cm; Resolution: 0.1 μS/cm.
Accuracy
High: ± 0.25 mS in the 0.1 - 200 mS range;
Low: ± 3 μS in the 0 - 100 μS range
Temperature Range
4-50° C
Temperature Accuracy
+/- 0.5° C
Temperature Probe
Thermistor
Pre-Calibration
0.10 M KCI Solution, 12.88 mS at 25.0° C
Sensor Microchip
EPROM, Stores Sensor ID, and
Calibration Factor.
Sensor Connector & Cables
Lockable and Waterproof.
SciPres:
Power
5 VDC provided by DINPres
®
Module
Fluid Connections
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Wetted Materials
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Sensor Type
Medical grade, silicone piezoresistive sensing
element with on-chip temperature compensation.
Sensor Isolation
Insoluble silicone dielectric gel isolates sensing
element from process solution. The gel is a non-
toxic, non-allergenic elastomeric system.
Pressure Range
-5 to 60 psi (-0.34 to 4.14 bar).