bit 2 = Start Span
bit 6 = Cal in Progress
bit 7 = Write Cal Data
221 R/W IEEE Float P1 Sensor Custom Pressure Offset (CPZ)
223 R/W IEEE Float P1 Sensor Custom Cal Factor (CCF)
235 R/W IEEE Float P1 Sensor Max Pressure
237 R/W 16 bit P1 Sensor MaxP, Day
238 R/W 16 bit P1 Sensor MaxP, Month
239 R/W 16 bit P1 Sensor MaxP, Year
241 R/W 16 bit P2 Sensor ID Type
0 = Luer
1 = 3/8" Barb
2 = 1/2" Barb
3 = 3/4" TC
4 =1" TC
100 = Cal Fixture
242 R/W 16 bit P2 Sensor ID Material
0 = Polypropylene
1 = Polysulfone
243 R/W 16 bit P2 Sensor ID Month
244 R/W 16 bit P2 Sensor ID Year
245 R/W 16 bit P2 Sensor ID Lot
246 R/W 16 bit P2 Sensor ID Unit Number
248 R/W IEEE Float P2 Sensor Pressure Offset (PZ)
250 R/W IEEE Float P2 Sensor Cal Factor (CF)
252 R/W IEEE Float P2 Sensor Cal Pressure (CP1)
254 R/W IEEE Float P2 Sensor Cal Factor 2 (CF2)
256 R/W IEEE Float P2 Sensor Cal Pressure 2 (CP2)
260 R/W 16 bit P2 Sensor Custom Cal Status
bit 0 = Invalid
bit 1 = Start Zero
bit 2 = Start Span
bit 6 = Cal in Progress
bit 7 = Write Cal Data
261 R/W IEEE Float P2 Sensor Custom Pressure Offset (CPZ)
263 R/W IEEE Float P2 Sensor Custom Cal Factor (CCF)
275 R/W IEEE Float P2 Sensor Max Pressure
277 R/W 16 bit P2 Sensor MaxP, Day
278 R/W 16 bit P2 Sensor MaxP, Month
279 R/W 16 bit P2 Sensor MaxP, Year
281 R/W 16 bit P3 Sensor ID Type
0 = Luer