5.8
Setup: Pump, continued:
SETUP: Pump provides configuration of global settings related to the pump. Generally, the default
values here do not need to be changed. Any changes required for a particular system will be made
at the factory, and would only need to be modified if a “Factory Reset” is performed.
KEYPAD BEEP: Determines if the buttons “beep” when pressed. Default = ON.
FACTORY RESET: Resets the system to the factory default settings for all menus.
Requires pressing “YES” to both the “Clear All Memory?” and “Are You Sure?” prompts.
ASCII FEEDBACK: Controls responses sent to a PC upon receipt of a remote command.
Default = ON.
EXTERNAL RUN-STOP: Controls action of the Footswitch connection, which is part of the External
I/O DB37 connector (pins 19 and 37). Choose Pulse for footswitch type control where the contacts
are closed and then opened. Choose Level for contact closure control where closed = Run and open
= Stop. Run key is disabled when this is set to Level except in Manual Mode. Default = Pulse.
POWER UP: Controls the action of the system upon power-up. Choose from Menu, Mode, or Run.
Menu = the system returns to the top of the Menu. Mode = the system returns to the last Mode it
was in. Run = the system returns to the last Mode it was in and starts the process.
PUMP TUBING: If Pump Head is set to Peristaltic, all nine standard tubing sizes can be chosen.
Used with Motor RPM setting to determine proper calibration curve. Default = 13.
PUMP HEAD: Choose between Peristaltic, RH1, RH0, RH00, Mag 201, Mag 122, Mag 120, and Mag
040. Used to access the proper calibration curve. Default = Peristaltic.
MOTOR RPM: Choose between 8, 160, 600 and 2400 rpm, matching the motor installed in the
system. This will be set at the factory. It will need to be reset after a Factory Reset if other than the
default. Default = 600.
MOTOR START: Controls how fast the motor starts. Hard is fast, Soft is slow. Default = Hard.
T T L 1 O N - O F F : C o n t r o l s a c t i o n o f T T L S w i t c h 1 , u s e d f o r M a s t e r / Slave control of another system
or device with TTL input control. Set to Yes, TTL 1 changes states from High to Low when the
system is told to run the motor. Set to No, it does not. Default = Yes.
SWITCH POLARITY: Controls the polarity of the TTL switches. Inverted = High when not activated,
Low when activated. Normal = Low when not activated, High when activated. Default = Inverted to
allow for proper Master/Slave control.
SWITCH CONFIGURATION: Sets the action of the TTL switches. Choose between Level and Pulse.
Default = Level to allow for proper Master/Slave control.