Place the sample bag or container onto the scale and underneath the dispensing tip. Press “RUN”
key
or the foot switch. The LabTec™ will show the following display:
This display is followed by:
Place the sample into the sample bag or container, only an approximate sample weight is needed.
The weight ratio will be applied to the actual sample weight as determined by the scale. Press “RUN”
key or the foot switch, the LabTec will start dispensing and display the progress:
SW stands for the actual sample weight and DW represents the delivered diluent weight. When the
dispensing cycle is completed, following displays are shown.
Remove your filled sample bag or container and initiate the second dispensing cycle.
REMOVING TARE WEIGHT
Please Wait
Press RUN When
SAMPLE IS IN VESSEL
SW= 10.00G
DW= 50.00 G
RUN
SW= 10.00G
DW= 50.00 G
DISPENSING
Completed
P
r
essRUNWhenRead
y
Note: If you have more than one WGT FACTOR, store one WGT FACTOR in Edit 1”, e.g. “9.00”,
while a second WGT FACTOR is stored in “Edit 2”, e.g. 99.00”. Up to ten different sets of
parameters can be stored in “Edit 1” through “Edit 10”.