EasyManua.ls Logo

Parkside PLDM A1 - Calibration

Parkside PLDM A1
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 GB
Calibration
NOTES:
o The thickness of the calibration plate[7] must be in the
range of 1.5mmto2.2mm.
o The initial display is 1.80mm. Press the TESTbutton[3]
to increase the reading on the display[2]. Press the
UNITbutton[4] to decrease the reading on the display.
o The reading should equal the thickness value of the
calibration plate. For example, if the thickness of the
calibration plate is 1.76mm, press the UNITbutton to
decrease the reading on the display to 1.76mm.
o The thickness of the calibration plate should be greater than
the layer thickness on coated panel.
1. Before a measurement in NFe or Fe test mode is made,
ensure that in this mode, the product has been calibrated
by using the naked substrate[8] of the object to be tested.
Alternatively a naked substrate can be used, if it is identical
to the substrate of the object to be tested.