Maintenance Schedule
14
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Routine Maintenance
Servicing and/or repair of the Generator should only be undertaken by a
TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated
condition.
SAFELY ISOLATED CONDITION
Definition: The unit is in a Safely Isolated Condition when it is disconnected from its
application, fully de-pressurised and isolated from the Electrical Supply. Directions for
isolating the generator are shown below.
Isolating the Generator:
a)Switch off the unit.
b)Disconnect the unit from the mains supply.
c)Ensure the internal pressure gauges (page 12) reads zero. (If gauge does not fall to
zero, loosen outlet fitting slightly to allow trapped gas to escape).
d)Disconnect from the application.
Maintenance Schedule
Note:
*Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 times, this is a more
cost effective solution, however a degree of technical expertise is required and can be time consuming.
Service kits are available for all routine maintenance. Please contact Peak Scientific for further details.
SERVICE INTERVAL COMPONENT PART NO.
AFD3000 FILTER ELEMENT (AFD) x 2 02-4335
RAC FILTER ELEMENT x 1 00-4425
ANNUAL SERVICE KIT ALL PARTS LISTED ABOVE 08-4435
2750 COMPRESSOR 08-1425
2688 COMPRESSOR 08-1420
COMPRESSOR TRAY ASSEMBLY (x1) 08-4649
2750 COMPRESSOR RE-FIT KIT (X1) 06-5529
2688 COMPRESSOR RE-FIT KIT (X1) 06-5542
on Request
1 YEAR
EVERY 6 MONTHS INLET FILTER (x4) 02-4187
COMPRESSOR (THE
LESSER) UNIT 6000 HRS
OR 18MONTHS
*ALTERNATIVE TO
COMPRESSOR ASSY.