Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully
inspect the unit for any signs of transit damage.
ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND
PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS
PURCHASED.
Following the un-packing instructions posted on the side of the crate. It will require 2 people to lift the
crate clear and to maneuver the generator onto the floor. Thereafter the generator can be moved to
its final location on the castors provided.
NOTE: - Included with the generator is a pack containing keys, manuals,
service filters & fittings. Be careful not to discard these with the packing.
Please save the product packaging for storage or future shipment of the
generator.
Following installation of the unit, care should be taken to ensure that the waste bottle is NOT a sealed
container, as this may cause possible pressure build-ups, explosion of the bottle, and consequential
damage to the unit. Any bottle used should have a form of exhaust hole, before being fitted to this unit.
Installation
Unpacking and Installation
Electrical Connection
Important Electrical Notice
This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED.
Before connecting the unit to the mains supply, please check the information on the serial plate.
The mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):- Green & Yellow OR Green
LIVE (L):- Brown Black
Neutral (N):- Blue White
Connect the generator to a single-phase supply using the power cord
provided.