EasyManua.ls Logo

PEDROLLO STEADYPRES V2.0 - Calibration of the Pressure Sensor

PEDROLLO STEADYPRES V2.0
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
- THEN assemble the PRESSURE SENSOR
- connect the wire communication with the power board
- connect the ribbon cable of the display and close the front cover
ATTENTION:
1. MAKE THE FACTORY PARAMETERS RESET BEFORE STARTING THE INVERTER (parameter
SET.F at page 11 of the Manual)
2. RESTORE MANUALLY THE SPECIFIC PARAMETERS OF THE SYSTEM
3. DO NOT RE-CALIBRATE THE PRESSURE SENSOR, IT IS FACTORY-CALIBRATED
CALIBRATION OF THE PRESSURE SENSOR
- You need an
auxiliary gauge
near STEADYPRES
-
bring the system (and STEADYPRES) pressure to zero (0 bar)
-
Start the calibration
of the pressure sensor according to the diagram below.
display action
P x.x
Switch the inverter off
display off
Switch the inverter on
STARTING
(for 10 sec.)
Press simultaneously the 4 buttons, within the 10 seconds of
STARTING.
ZERO
Check on the gauge that the pressure of the system is ZERO
Press the button SET/TEST to confirm the zero pressure
SP 5.0
Start the pump by pressing START button; the pump starts at
the minimum frequency
SP 5.0
Increase the frequency of the pump by pressing the button
until you reach the required pressure of 5 bar
If the pump is not able to reach 5,0 bar, reduce the required pressure
by pressing the button
(e.g. to 4 bar)
SP 4.0
Check on the gauge that the pressure of the system is that required by
the inverter
(e
.g
. 4 bar)
Press the button ENTER to confirm
F hi
Press again the button ENTER
P_3.5
The inverter re-starts in automatic mode, the sensors calibrated.

Table of Contents