8-20 Command Reference
:CALCulate :LIMit :LOWer :STATe
8 <Boolean>
Check Against Lower Limit
Selects if the measured value should be checked against the lower limit.
Parameters <Boolean> = ( 1/ON | 0/OFF )
Returned format: 1| 0 ¿
*RST condition: 0
Complies with standards: SCPI 1991.0 confirmed.
:CALCulate :LIMit :UPPer
8 «<Decimal data>|MAX|MIN»
Set Upper Limit
Sets the value of the ‘Upper Limit’, i.e., the highest measurement result allowed
before the counter generates a 1 that can be read with :CALCu
-
late:LIMit:FAIL?, or by reading the corresponding status byte.
Parameters
Range: –9.9*10
+37
to +9.9*10
+37
Returned format: <Decimal data>¿
*RST condition: 0
Complies with standards: SCPI 1991.0, confirmed.