EasyManua.ls Logo

Pendulum CNT-91 - Configure:Array:<Measuring; Function

Default Icon
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-29 Command Reference
:CONFigure :ARRay :<Measuring Function>
8 (<array size>)[,<parameters> [,(<channels>)]]
Configure the counter for an array of measurements
The :CONFigure:ARRay command differs from the :CONFigure command in
that it sets up the counter to perform the number of measurements you choose in
the <array size>.
To perform the selected function, you must trigger the counter with the :READ:AR
-
Ray? or :INITiate;:FETCh:ARRay? queries.
Parameters <array size> sets the number of measurements in the array. See table be
-
low.
<Measuring Function>, <Parameters>, and <Channels> are defined on page 8-54.
Example:
SEND®
:CONF:ARR:PER 8 (7),5E–3,1E–6,(@4)
This example sets up the counter to make seven period measurements. The ex
-
pected result is 5 ms, and the required resolution is 1 ms. The EXT ARM input is
the measuring input.
To make the measurements and fetch the seven measurement results:
SEND®
:READ:ARR? 8 7
READ¬
5.23421E-3,5.12311E-3,5.87526E-3, 8
5.50345E-3,5.33901E-3,5.25501E-3, 8 5.03571E-3
Function
Array Size
Interpolator Calibration ON Interpolator Calibration OFF
Freq, Per &
most other
functions
CNT-90 CNT-91 CNT-90 CNT-91
375 k 1.9 M 750 k 3.8 M
Smart Freq
&Per
30 k
Volt (Max,
Min, PTP)
10 k
Volt Ratio 30 k
Totalize NA 3.8 M NA 3.8 M
Note: To find out the maximum number in your particular case, do the following:
1. Make all settings not affected by CONFigure
2. Send :CONF:ARR:<meas. func> MAX,(@<channel>)
3. Send Query :TRIG:COUN? to see the maximum number of samples for this
measurement
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-91

Related product manuals