Using the Active Method
3-28
The Carrier Velocity Program window appears:
3. Touch each individual field, ranging from the Initial field to the step 4 field, and
use up and down arrow or keypad
buttons to enter a velocity,
equilibration time and rate.
NOTE: The rate for the range is 0 (end) to 99. A rate of 999 is a go to setting
CAUTION
The go to mode of programming is not recommended. Setting a
program rate will be more reproducible.
4. Touch Close.