Clarus 600/680 User’s Guide
3-17
Oven Temperature Programming Mode
In this mode the POC or PSS injector is configured as Oven. The injector temperature is
set to the oven temperature plus 5 °C. This is the default mode and the easiest mode to
operate in. For example, if the oven program is as follows:
The initial injector temperature will be 55 °C. When the GC run begins, the injector will
hold this temperature for 2 minutes, then it will ramp up to 245
°C at 5 °C/min and hold
for 10 min.
Figure 3-3. PSS oven mode temperature program.
NOTE: Opening the oven door turns off the oven and the injector heater. Also, if you reset the
oven temperature during the run to the next level, the injector temperature will follow.