CAUTION
If you are using pressure pulse injection, you must set the split flow on just
prior to when the inlet pressure begins to drop to the operating setpoint.
For example, if the inlet pressure is programmed to start decreasing one
minute after injection, use a split flow event time of 0.98 minutes.
At the end of a run, the split flow resets to the original setting. Setting a Value of
50 mL/min minimizes contamination by keeping the splitter on during cool-
down.
NOTE: The timed event that you set up, or valve settings that you make, become part of the
method. The timed event is automatically saved if you are editing the Active Method.
NOTE: When operating in the splitless mode, the recommended measured flow out of the vent is
30 - 50 mL/min. when the split vent is ON.