Setting Up the Detectors
4-16
WARNING
The FID outlet is hot! To prevent injury, never place your
hand over the FID outlet.
2. Lift open the detector cover.
3. Cut the piece of tubing (P/N 0250-6490) to a length of 4-inches and firmly
insert one end of the tubing into the top of the FID.
4. Route the end of the tubing through the hole in the detector cover as you close
the detector cover.
5. Connect one end of a length of vent tubing to this piece of tubing and place the
other end in a vent hood.