Clarus 600/680 GC User’s Guide
2-19
Configuration Options
The following screens show the configuration choices and how they correspond
to the entry fields in the active method.
Configuration Screen Active Method
Temperature Program
You can set injector (inlet) temperature programs by selecting, the program
button
on the active method screen is active.
Inlet
In the inlet programmed mode, the injector is first heated to the
injector temperature that is entered in the method. The injector
then follows the injector temperature program that you defined.
Oven
The oven programmed mode is the default mode of operation. In
this mode, the injector setpoint equals the oven setpoint plus 5 °C.