Using the Active Method
3-26
The Carrier Pressure Program window appears:
3. Touch to highlight each individual field, ranging from the Initial field to the
step 4 field, and use up and down arrow or keypad
buttons to enter
a pressure, HOLD time and HOLD rate.
NOTE: The rate for the range is 0 (end) to 99. A rate of 999 is a go to setting.
CAUTION
The go to mode of programming is not recommended. Setting a
program rate will be more reproducible.
4. Touch Close.