Template-Company IQ / OQ
Template-SN Page 1 of 29
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Liquid Chromatograph
Flexar Analytical / FX-10 UHPLC Pumps
Installation Qualification (IQ) and Operational Qualification (OQ)
Company Name:
Template-Company
Address:
Template-Address
Location, Room:
Serial Number or System Name:
Template-SN
Protocol Number:
Template-Protocol
Issue Date:
27 May 2009
Date Tested:
Valid if tested within 1 year of Issue Date
Recertification Period
Recommended at 12 Months
Recertification Due Date:
Release History
Part Number Release Publication Date
09234069 D June 2009
Any comments about the documentation for this product should be addressed to:
User Assistance PerkinElmer Global Technical Support
PerkinElmer LAS (UK) Ltd M/S 215
Chalfont Road 710 Bridgeport Avenue
Seer Green Shelton
Beaconsfield Connecticut 06484-4794
Bucks HP9 2FX U.S.A.
United Kingdom
Question and Answer IconNeed help?

Do you have a question about the PerkinElmer FX-10 UHPLC and is the answer not in the manual?

PerkinElmer FX-10 UHPLC Specifications

General IconGeneral
BrandPerkinElmer
ModelFX-10 UHPLC
CategoryMeasuring Instruments
LanguageEnglish

Summary

Protocol Documentation

Error Code Abbreviation, Definition, and Making Correction

Approval Page

Part I - Installation Qualification

Installation Qualification Checklist

Checklist for verifying installation components, setup, software, and system familiarity.

Installation Qualification Verification

Confirms successful installation and availability of supporting documents.

Part II - Operational Qualification

Parameter Testing

Details specific tests to verify instrument performance parameters.

Pressure Pulsations Test

Verifies output flow fluctuations by measuring pressure variations.

Flow Accuracy (1 mL/min)

Tests flow accuracy for analytical pumps at 1 mL/min using gravimetric method.

Flow Accuracy (3 mL/min)

Tests flow accuracy for analytical pumps at 3 mL/min using gravimetric method.

FX-10 Flow Accuracy (0.3 mL/min)

Tests flow accuracy for FX-10 pumps at 0.3 mL/min using gravimetric method.

FX-10 Flow Accuracy (0.9 mL/min)

Tests flow accuracy for FX-10 pumps at 0.9 mL/min using gravimetric method.

Composition Accuracy Test

Tests proportioning valve accuracy by measuring absorbance of solvent mixtures.